A. load issues, Load issues, Appendix a – Fluke Biomedical RF303 User Manual

Page 65

Advertising
background image

A-1

Appendix A

Load Issues

Load Issues

The load resistors typically used in ESU analyzers are not ideal; they have

some reactive components that are frequency dependent. The RF303 Analyzer

derives applied power by measuring the voltage across the set load and

calculating the power (V2/R). Most other ESU analyzers on the market derive

the applied power by measuring the current flowing through a set load and

calculating the power (I2*R).

At fundamental frequencies below 500 kHz and regardless of the load setting,

the two methods of measurement are comparable. Above 500 kHz, and at the

extremes of the loads, the readings displayed by the two methods differ on

opposite sides of the expected value.

As an example, when testing the Conmed Excalibur Electrosurgical Unit in the

monopolar output with the Analyzer load set to 50 ohms, the set value on the

Conmed correlates with the displayed value on the Analyzer. When in the

bipolar mode, the Analyzer displays higher than expected values – up to 35%

higher. The same test performed on some current measuring analyzers

produces lower than expected values. This discrepancy is caused by the

difference in fundamental frequencies between the monopolar and bipolar

modes. The Conmed operates at 500 kHz in monopolar mode and 1 MHz in

bipolar mode. When comparing readings measured with a V2/R device to

those measured with an I2*R device, the total measurement difference is

usually larger than 35%, because of different methods of deriving power.

This difference does not mean that the Analyzer is malfunctioning or is in

error. Rather, it reflects the different results that the two measurement

techniques produce when the load deviates from the nominal value used in the

power calculations.

Note

Most ESU manufacturers use the current measuring technique to
calibrate production units.

Advertising