Fluke Biomedical RF303 User Manual
Page 36
Advertising
RF303
Operators Manual
2-12
5. Reinstall the fuse carrier into the power inlet module, ensuring that the
arrows on the ends of the fuse carriers point up.
6. Close the fuse cover until it snaps into place.
7. Reconnect the power cord, and power-up the Analyzer, if desired.
If you have replaced these fuses and the Analyzer still does not work properly,
contact the Fluke Biomedical Technical Assistance Center at 800-648-7952.