General specifications – Fluke Biomedical RF303 User Manual

Page 21

Advertising
background image

Introduction and Specifications

General

Specifications

1

1-11

Front-Panel Controls / Pushbuttons

Measurement Select (1)
Load Select:

Increment test load (+) one step
Decrement test load (-) one step

Designations:

Generator output-active (1)
Generator output-dispersive (2)
Auxiliary HF leakage load (2)

Connector type ................................................... 4-mm (0.160”) diameter safety sockets
Input voltage limit................................................ 10,000 V peak
Input current limit ................................................ 3 amperes rms
Installation category............................................ II

Side Input connection

Designation......................................................... Signal reference

Oscilloscope Output

Transformer coupled output
Scale Factor........................................................ uncalibrated
Connector type ................................................... BNC

Calibration Period

Calibration recommended every 12 months.

General Specifications

Temperature Range

Operating: ........................................................... 15 to 35 °C
Storage ............................................................... 0 to 50 °C

Humidity Range

90 % non-condensing

Altitude

To 2,000 meters

Ventilation

Internal fan with variable speed control
Over temperature detector
Magnetic tachometer sensor to detect blocked fan motor