Fluke Biomedical RF303 User Manual

Page 53

Advertising
background image

Operation

Testing

the

ESU

3

3-17

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz07.eps

To measure the ESU’s signal HF leakage:

1. If the

mA

indicator lamp on the panel is lit, go to Step 2. If it is not lit,

press the

MODE SELECT

(

ENT

) button to select

mA

and go to Step 2.

2. Activate the ESU and view the signal HF leakage measurement on the

display. The measurement will be in milliamperes. Here is an example.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz18.eps

To end HF leakage test number three:

Deactivate the ESU to end the HF leakage test.

W

Caution

To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer’s duty cycle, which is 30
seconds ON during any continuous one-minute period.

Advertising