Fluke Biomedical RF303 User Manual
Page 59
Operation
Remote Operation via the RS-232 Function
3
3-23
Table 3-2. Available Duplex Mode Commands
Command Description
Parameters
Required
Return Value
(Response)
GOTOREMOTE
Places the Analyzer in
the duplex remote
mode.
None *
GOTOLOCAL
Returns the Analyzer to
normal (local mode)
operation / exits duplex
remote mode.
None
*
IDENT
Identifies the Analyzer
model and firmware
version number.
None
Model Analyzer name
and the version of the
firmware.
Example: MODEL
Analyzer VERSION
X.XX
Where X.XX is the
version number.
VER
Identifies the firmware
version number.
None
Firmware version
number.
Example: VERSION
X.XX
Where: X.XX is the
version number.
SETMODE:{parm}
Changes the format of
the Analyzer response
to the RDMETER
command. Separate the
command from the
desired mode
(parameter) using a
colon. Refer to
RDMETER below.
Example:
SETMODE:OUTPUT
OUTPUT (for
measuring
generator
output).
RFLKG (for
measuring
HF leakage).
RARF (for
conducting
CQM test).
*