Fluke Biomedical RF303 User Manual

Page 59

Advertising
background image

Operation

Remote Operation via the RS-232 Function

3

3-23

Table 3-2. Available Duplex Mode Commands

Command Description

Parameters

Required

Return Value

(Response)

GOTOREMOTE

Places the Analyzer in
the duplex remote
mode.

None *

GOTOLOCAL

Returns the Analyzer to
normal (local mode)
operation / exits duplex
remote mode.

None

*

IDENT

Identifies the Analyzer
model and firmware
version number.

None

Model Analyzer name
and the version of the
firmware.

Example: MODEL
Analyzer VERSION
X.XX

Where X.XX is the
version number.

VER

Identifies the firmware
version number.

None

Firmware version
number.

Example: VERSION
X.XX

Where: X.XX is the
version number.

SETMODE:{parm}

Changes the format of
the Analyzer response
to the RDMETER
command. Separate the
command from the
desired mode
(parameter) using a
colon. Refer to
RDMETER below.

Example:
SETMODE:OUTPUT

OUTPUT (for
measuring
generator
output).

RFLKG (for
measuring
HF leakage).

RARF (for
conducting
CQM test).

*