Operation

Testing

the

ESU

3

3-9

To observe the ESU’s visual or audible alarm:

1. With the 50-ohm test load attached to the ESU CQM (REM) input, it

should sense this level of resistance in the "pass zone." Depending upon
the ESU device manufacturer, this pass zone can extend from 135 to 250
ohms. Test load settings over the pass zone activate the CQM alarm.

Note

Refer to the technical support/service manual of the ESU you are
testing for the recommended check-out procedure for the ESU CQM
feature.

2. Increase the test load from 50 ohms until the CQM alarm sounds. The

value at that point is the CQM alarm resistance.

DISPERSIVE

ACTIVE

ESU

Basic CQM Check

(Contact Quality Monitor)

RF303

X No connection

ESU CQM

BLU

YEL

Main

Test

Load

Supply
Mains

Earth

CQM

exz13.eps

Figure 3-2. Basic CQM Check