Fluke Biomedical RF303 User Manual
Page 42
![background image](/files/785354/content/doc042.png)
RF303
Operators Manual
3-6
To measure the ESU’s power output:
1. Activate the ESU.
2. If the
WATTS
indicator lamp on the panel is lit, go to Step 3. If it is not
lit, press the
MODE SELECT
(
ENT
) button to select
WATTS
and go to
Step 3.
3. View the power measurement on the display. Here is an example:
WATTS
mA
OHMS
ENT
MODE SELECT
OHMS SELECT
exz04.eps
To measure the ESU’s current output:
1. While the ESU is still activated, press the
MODE SELECT
(
ENT
) button
until the milliamperes (
mA
) indicator lamp lights.
2. View the ESU’s current output (in mA). Below is an example.
WATTS
mA
OHMS
ENT
MODE SELECT
OHMS SELECT
exz05.eps
To end the generator output test:
Deactivate the ESU.
W
Caution
To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer’s duty cycle, which is 30
seconds ON during any continuous one-minute period.