To measure the ESU’s power output:

1. Activate the ESU.

2. If the

WATTS

indicator lamp on the panel is lit, go to Step 3. If it is not

lit, press the

MODE SELECT

(

ENT

) button to select

WATTS

and go to

Step 3.

3. View the power measurement on the display. Here is an example:

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz04.eps

To measure the ESU’s current output:

1. While the ESU is still activated, press the

MODE SELECT

(

ENT

) button

until the milliamperes (

mA

) indicator lamp lights.

2. View the ESU’s current output (in mA). Below is an example.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz05.eps

To end the generator output test:

Deactivate the ESU.

W

Caution

To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer’s duty cycle, which is 30
seconds ON during any continuous one-minute period.