Fluke Biomedical RF303 User Manual

Page 20

Advertising
background image

RF303

Operators Manual

1-10

Test Load Section

Main Test Load

Selections........................................................15
Selection range ...............................................50 to 750 e
Step size..........................................................50 e
Duty Cycle .......................................................50 % @ 400 W (maximum 30 seconds

ON during any one-minute period)

Resonance impedance variation ..................... ± 0.5 dB maximum (<10 MHz)

Accuracy (DC to 500 kHz)...................................± 4 % of selected value measured at

calibration to: ± 1 % (across the entire
operating temperature range)

Auxiliary Leakage Test Load Fixed:

200 e
Accuracy..........................................................± 4 %
Power rating ....................................................225 W

Input Capacitance (nominal)

Active to dispersive .........................................30 pF
Active or Dispersive to Earth ground...............40 pF

Battery

Type: Sealed lead-acid
Voltage ................................................................12 volts nominal
Capacity ..............................................................2.2 A H
Field serviceable .................................................No
Typical time between recharges..........................2-hour minimum
Battery cycles ......................................................200
Recharging ..........................................................Instrument has internal, automatic

charger that is activated when unit is
plugged into wall with power cord. No
external charger required.

Auxiliary Contact Quality Monitor

Testing Feature ...................................................The main test load section is used to

perform a simple Auxiliary Contact
Quality Monitor Testing Feature
(CQM) operational check.

Display

Type ....................................................................LCD, 7-segment
Display size .........................................................Four full digits
Overall display size .............................................2.0” x 0.75”

Advertising