Operation

Testing

the

ESU

3

3-11

3. If operating under battery power, attach a test lead from the set (Part #

1903307) from the battery ground on the side panel to the ground earth
reference.

To select a 200-ohm test load resistance setting:

1. Press the

OHMS SELECT

increase (+) or decrease (-) buttons as needed

until the test load value 200 appears in the display. After approximately
one second, the display returns to a blank screen. Below is an example of
a test load value as it appears in the display.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz07.eps

To measure the ESU’s signal HF leakage:

1. If the

mA

indicator lamp on the panel is lit, go to Step 2. If it is not lit,

press the

MODE SELECT

(

ENT

) button to select

mA

and go to Step 2.

2. Activate the ESU and view the signal HF leakage measurement in

milliamperes on the display. Here is an example.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz18.eps

To end HF leakage test number one:

Deactivate the ESU to end the HF leakage test.