Test procedures – Fluke Biomedical RF303 User Manual

Page 44

Advertising
background image

RF303

Operators Manual

3-8

Test Procedures

The procedures involved with this test include:

connecting the ESU neutral/dispersive electrode to the Analyzer

selecting a 50-ohm test load resistance

increasing load resistance on the Analyzer to observe the ESU’s

visual or audio alarm

To connect the ESU’s neutral/dispersive electrode to the Analyzer:

1. Connect one of the two color-coded banana plugs (test lead Part #

2772180) to the Analyzer’s yellow jack active.

2. Connect the other banana plug to the Analyzer’s left blue jack dispersive.

3. Connect the 2PIN CQM (REM) connector on the other end of the test lead

to the neutral/dispersive electrode of the ESU.

To select a 50-ohm test load resistance setting:

1. Press the

OHMS SELECT

increase (+) or decrease (-) buttons as needed

until the test load value 50 appears in the display. After approximately one
(1) second, the display returns to a blank screen. Below is an example of a
test load value as it appears in the display.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz12.eps

Advertising