Fluke Biomedical RF303 User Manual

Page 41

Advertising
background image

Operation

Testing

the

ESU

3

3-5

2. Connect the ESU Neutral/Dispersive electrode to the left blue jack

dispersive, using the test lead, Part # 2772171.

3. Continue to load selection.

To connect the ESU to the Analyzer for a bipolar output test:

1. Connect the ESU Bipolar-1 electrode to the yellow jack (marked) active,

using one of the test leads from the set, Part # 1903307.

2. Connect the ESU Bipolar-2 electrode to the left blue jack dispersive, using

the other test lead from the set, Part # 1903307.

3. Continue to load selection.

To select the desired test load resistance:

1. Press the

OHMS SELECT

increase (+) or decrease (-) buttons until the

test load value you want appears in the display. After approximately one
second, the display returns to a blank screen. Below is an example of a test
load value as it appears in the display.

WATTS

mA

OHMS

ENT

MODE SELECT

OHMS SELECT

exz03.eps

Note

The default (or initial power-up) test load selection is 300 ohms.

Advertising