Fluke Biomedical RF303 User Manual
Page 35

Installation, Setup, and Maintenance
Maintenance
2
2-11
exz16.eps
Figure 2-1. Fuse Cover
W
X
Warning
To avoid severe electrical shock, disconnect the power
source before replacing fuses.
1. Make sure that the power switch is in the OFF position (
O)
and disconnect
the power cord. The Analyzer is now safely powered down.
2. Pry the left side of the fuse cover up from the power inlet assembly, using
a small flat-blade screwdriver. The fuse cover latches on the right side and
remains attached to the Analyzer.
3. Remove a fuse carrier with your fingers, lifting it from the power inlet
housing.
Note
To avoid damage to the Analyzer, do not use any sharp instrument to
lift a fuse carrier.
4. Replace the fuse in the carrier with one of the same amperage and voltage
ratings.