Appendix B

Interpretation of Fluctuating

Readings

Interpretation of Fluctuating Readings

The RF303 Analyzer incorporates an accurate digital measurement system that

operates in three modes. The first measurement mode (default mode) utilizes a

relatively short sampling time and does not filter or average ESU output. The

two selectable measurement modes utilize longer sampling times that average

ESU output. When using the default measurement mode, you may observe

some ESU output readings that fluctuate plus or minus 10% or more,

depending on the unit under test. This is normal operation and is not indicative

of a problem with the Analyzer.

When fluctuating readings are observed, take note and determine if this is

normal behavior for the ESU under test, or if this behavior is a sign of a

problem.

Older ESUs may have unstable output due to old technology or

design. This is typical and is acceptable, based on the manufacturer’s

limits.

Many newer-generation ESUs utilize instantaneous feedback loops,

which constantly adjust the output and can cause an oscillating effect.

This result is also satisfactory and is considered normal for these

devices.

In some cases, fluctuating ESU power output is evidence of a

problem. Fluctuating output on some ESUs may indicate the

weakening of output from power transistors or other ESU ailments

and is not acceptable.

If an ESU normally has fluctuating output, then operating the Analyzer in one

of the Signal Averaging Modes (SAM) may be desirable. SAM significantly

reduces fluctuating readings on the display of the Analyzer making it easy to

read an average value.