RF303

Operators Manual

3-24

Table 3-2. Available Duplex Mode Commands (cont.)

Command Description

Parameters

Required

Return Value

(Response)

RDMODE

Prompts the Analyzer to

transmit the currently

selected mode.

None NONE

OUTPUT

RFLKG

RARF

EXITMODE

Prompts the Analyzer to

exit the currently

selected mode.

Note: You must exit the

current mode before

entering another.

None *

SETLOAD:{parm}

Sets the measurement

test load resistance from

50-750 ohms in 50 ohm

increments. Not

available in RFLKG

mode.

Values of 50-

750 in

increments of

50

*

RDLOAD

Resistance value of the

selected test load

appears in the display.

None RRRR

(ohms)

Where RRRR is the

load value in ohms.

Formatted to four

digits with leading

spaces as required.

RDPOWER Prompts

an

Analyzer

power measurement in

watts. Not available in

RFLKG and RARF

modes.

None PPP.P

(w)