Fluke Biomedical DataSim 6100 User Manual

Page 19

Advertising
background image

Introduction and Specifications

Instrument

Familiarity

1

1-7

Table 1-2. Simulator Front Panel Controls and Indicators

Label Component

Description

A

Power Indicator

Illuminates when power is ON

B

Power Switch

Switches power on and off

This battery operated unit is
rechargeable with the supplied charger.

C

ECG Snap
Connectors

Outputs low level 12 lead ECG

Attach standard patient ECG cable lead
wires to these snap connectors to
display the simulated ECG waveforms
on the monitor screen.

To minimize 60-cycle artifact, use a
reference electrode connection. For
example, most five-lead ECG diagnostic
recorders require that a Simulator REF
signal is connected to the RL input of
the patient cable. The Simulator can
also output the ECG in a modified chest
lead configuration (MCL2) used for
bedside monitoring. Simply attach the
LL (+) red, RA (-) white, and LA (ref)
patient cable snaps to the patient
Simulator. Refer to the operators
manual of the ECG monitor for hook-up
instructions.

D

Pressure Output
Connectors

Output for arterial, PA, and RA
pressures

This output simulates the electrical
output of the BP transducer that would
be used with the patient monitor. The
Simulator generates signals that are
compatible with either 5 or 40-microvolt
transducers.