Interactive iabp personality modules – Fluke Biomedical DataSim 6100 User Manual

Page 52

Advertising
background image

DataSim 6100

Operators Manual

2-16

Table 2-6. Personality Modules (cont.)

MDE Ref #

Part #

Description

Additional Detail

6070-29B 2244678

Interactive
IABP/Datascope

System 90

6070-30B 2244684

Interactive
IABP/Aries

Early Inflation
Late Inflation
Early Deflation
Late Deflation
Proper Timing

6070-37B 2244706

Normal/Disease
Heart

Interactive IABP Personality Modules

The Interactive Intra-Aortic Balloon Pump Personality Modules are used with
the Simulator and Kontron, Datascope System 80, Datascope System 90, and
Aries monitors. Invasive arterial BP and sync timing cables are included with
each module.

To generate the special waveforms:

1. Determine the need for a balloon according to the monitor used.

Datascope 90: It is essential that a balloon (or suitable replacement)
be connected to the

Balloon Output

connector for proper operation

of the Datascope 90 pump.

Datascope 80, Kontron Aries: It may be necessary to defeat the alarm
by connecting a balloon or suitable replacement to the

Balloon

Output

connector.

2. Connect Datascope/Kontron/Aries ECG cable to the Simulator

ECG

snap

connectors.

3. Insert the Interactive IABP module into the Simulator

Personality

Module

connector.

4. Power on the Simulator.

5. Connect the following cables to their respective monitors: