The inter-connect cable is not supplying an excitation reference
voltage required by the Simulator.

The inter-connect cable is wired to the wrong signal output pin.

Respiration artifact on the pressures modulates the pressure values in
sync with the respiration cycle. This artifact significantly alters the
systolic/diastolic values.

5. To remove respiration artifact, select

SPEC FUNC 67

(

PRES RESP

)

and decrease to the minimum level.

Personality Modules (Optional Special
Functions)

Optional Personality Modules increase the number of Special Functions
available in the Simulator. These can be plugged into the front panel connector
and provide the following scenarios:

Interactive IABP – helps students practice setting inflation/deflation
timings. The Simulator can be interfaced with all Kontron, Datascope, and
Aries balloon pumps. The Simulator responds to the inflation/deflation
timing of the balloon pump by changing the timing and shape of
augmented arterial pressures.

Cardiac Output – simulates normal and abnormal thermal dilution curves
by connecting the Simulator to any cardiac output computer that accepts
the Edwards 3-pin catheters.

PALS Training – an optional Pediatric ECG Personality Module used with
the DEFIB training feature during Pediatric (PALS) training workshops.

Note

Power off the Simulator before inserting or removing a Personality
Module.

Each module has a function menu listing the functions and code numbers
within the module’s memory. These functions are assigned code numbers
between 89 and 99. Available Personality Modules are listed in Table 2-6.