Table 1-5. Pendant Keypad Codes and Functions (cont.)

Code Function

Description

34 SENSE

and

CAPTURE FAIL

Frequent loss of capture, low amplitude
escape beat, sensing failure of escape
beat

35 FAILURE

TO

CAPTURE

No capture; pacer artifact amplitude is 8
mV with 1 msec duration.

MISC ECG

38 CARDIAC

FAILURE

Intermittent PVCs and couplet followed
by ventricular tachycardia, followed by
ventricular fibrillation, followed by
asystole

39

ST ELEVATION

Regular rhythm with ST Elevation at 2
mm 80 msec after J point

40

ST DEPRESSION

Regular rhythm with ST Depression at 3
mm 80 msec after J point

41

AGONAL

No P waves; wide complexes with
periods of asystole

42 VENTRICULAR

ASYSTOLE

No QRS, regular atrial rhythm

43

CONVERSION

Vfib with CPR followed by Vfib without
CPR for 2 seconds prior to defib: Defib
artifact, Brady rhythm with CPR, Brady
rhythm without CPR, followed by NSR

ECG ARTIFACT

46

DEFIBRILLATION

Positive and negative baseline
saturation; artifact is automatically
followed by pre-selected rhythm (see
#59 below).

47 60

CYCLE

ARTIFACT

Artifact amplitudes 0.1, 0.4, and 0.7 mV
p-p; use

]

and

[

keys or eliminate with

STD key.

48 ECG/RESP

ARTIFACT

Artifact amplitudes 0.4, 0.6 and 1.0 mV
peak to peak; use

]

and

[

keys or

eliminate with STD key.