Introduction and Specifications

Instrument

Familiarity

1

1-11

27

. Functions can be selected in any order to simulate virtually any patient

condition sequence.

After the code is input, the following prompt displays:

], [ or STD

At this prompt, the user can adjust the value of the parameter as desired:

]

- Increases parameter value; momentarily displays at max“ when

maximum value is reached

[

- Decreases parameter value; momentarily displays at min when

minimum value is reached

STD - Resets parameter value to the standard value

Some of the Special Functions are actually programmed sequences of rhythms
or events. Table 1-5 describes the codes and functions.

Table 1-5. Pendant Keypad Codes and Functions

Code Function

Description

00

NORMAL SINUS

Provides a convenient way to return to
normal sinus rhythm at a rate of 70 BPM

01 SINUS

TACHYCARDIA

Rapid regular rhythm with normal P
wave and heart rate of 140 BPM

02 SINUS

BRADYCARDIA

Slow regular rhythm with normal P wave
and heart rate of 35 BPM

03

ASYSTOLE

No beats but very small irregular signal
remains

04 ATRIAL

FLUTTER

Regular ventricular rhythm; atrial flutter
with 4:1 conduction

05 ATRIAL

FIBRILLATION

Rapid irregular atrial signal with no real
P waves; irregular ventricular rate

06 2nd

DEGREE

AV

BLOCK I

Wenckebach; irregular rhythm with
normal P waves; PR interval lengthens
progressively until a dropped beat
occurs and cycle repeats.