Respiration – Fluke Biomedical DataSim 6100 User Manual

Page 44

Advertising
background image

DataSim 6100

Operators Manual

2-8

To enable an artifact:

1. Using the keypad, press

SPEC FUNC

, followed by the numbers

47, 48,

or

49

. For example, using the number

47

, the keypad briefly displays:

60 CYCLE

followed by the prompt:

], [ or STD

2. Press

]

to increase or

STD

to reset to standard level. The

[

capability is

not available.

Table 2-5 shows the effect of artifact modification on the parameters.

Table 2-5. Artifact Modifier Effects

FUNC TYPE STD +1 +2 +3

47

ECG 60 Hz

0

.1

.4

.7 mV

48

ECG RESP

0

.4

.6

1.0 mV

49

ECG MUSCLE

0

.4

.6

1.0 mV

Respiration

The Simulator provides a varying thoracic impedance signal via all 10 ECG
electrode output snaps. Several respiration rates from apnea to 80 BPM can be
selected. After a respiration rate has been selected, the respiration rate remains
fixed during subsequent rhythm selections or until a new respiration rate is
selected.

To vary the amplitude of the thoracic impedance signal (R):

1. Using the keypad, press

SPEC FUNC

, followed by the numbers

54

. The

keypad briefly displays:

SIZE