Modifier/cal mode – Fluke Biomedical DataSim 6100 User Manual

Page 30

Advertising
background image

DataSim 6100

Operators Manual

1-18

Modifier/Cal Mode

The bottom row of the keypad has three dual function keys:

CPLT/up arrow

V-TAC/down arrow

V-FIB/STD

Pressing one of these keys without first pressing the

SPEC FUNC

key puts

the Simulator into the Modifier/Cal mode. This mode can also be used in
conjunction with the functions listed in Table 1-6.

Table 1-6. Functions Used in the Modifier/Cal Mode

Code Function

53 Rate

Adjust

54 Size

78

Heart Rate Cal 70 BPM

82

ECG Size Cal

84

Resp Size Cal

47

60 Cycle Artifact

48 ECG/Resp

Artifact

49 ECG/Muscle

Artifact

After the Modifier/Cal mode is invoked and one of the functions is selected,
the user can press one of these dual function keys to increase, decrease, or
reset the standard value for the selected parameter.

Pressing any of the remaining keys exits the Modifier/Cal function, retains the
parameter level, and simulates the rhythm selected (NSR, Bradycardia, etc.)