Artifact simulation – Fluke Biomedical DataSim 6100 User Manual

Page 43

Advertising
background image

Operation

Artifact

Simulation

2

2-7

Table 2-4. Heart Rates Obtained via Rate Adjust Special Function

RHYTHM -40%

-30%

-20%

-10%

STD

+10%

+20%

+30%

Bradycardia 21 24 28 31 35 38 42 45

Ventricular 21 24 28 31 35 38 42 45

Junctional 24 28 32 36 40 44 48 52

Normal
Sinus

42 49 56 63 70 77 84 91

Sinus Tach

84

98

112

126

140

154

168

182

Atrial Tach

90

105

120

135

150

165

180

195

Atrial

Fib

60 70 80 90

100

110 120 130

Note

For performance check applications, the Heart Rate Calibration,
Special Function 78 (ECG RATE), provides calibrated heart rates
from 30 to 300 BPM, See Table 2-3.

To adjust the heart rate:

1. Using the keypad, press

SPEC FUNC

, followed by the number

53

. The

keypad briefly displays:

RATE

followed by the prompt:

], [ or STD

2. Press ] to increase or [

to decrease the value, or press

STD

to reset to the

standard level.

Artifact Simulation

Various types of artifact can be superimposed on the ECG. The type and level
of artifact is selected via the Special Functions 47, 48, and 49.