Fluke Biomedical DataSim 6100 User Manual

Page 46

Advertising
background image

DataSim 6100

Operators Manual

2-10

5. To enable the automatic defib conversion feature and select the post defib

rhythm:

a. Press and release the

SPEC FUNC

key

b. Enter code number

59

to initiate the

DEFIB CONVERSION

Rhythm Select

function. The keypad displays:

SEL RTHM

c. Select the desired post-defib rhythm by pressing a Direct Function

key, or by pressing the

SPEC FUNC

key followed by the two-digit

code number of the desired rhythm.

Note

If a post defib rhythm is not programmed, the defib artifact is
followed by asystole until another rhythm is manually selected.

6. To initiate the defibrillation sequence:

Defibs equipped with apex/sternum paddles: Firmly place the paddles
on the electrodes located on the manikin chest skin or torso cover.

Defibs equipped with disposable electrodes: Plug in adapters and
connect cables from defib to adapter snaps.

7. Charge the defib to 25 joules for Arrhythmia Anne or up to 400 joules for

Chris Clean.

8. Press the

SIMULATE DEFIB

button.

W

Caution

To avoid

damage to the Simulator or adverse affects on its

performance, refer to

the Arrhythmia Anne and Chris Clean

instruction manuals for the proper procedures and use of
the defibrillator on the manikin. Under no circumstances
should a defibrillator be discharged directly into the
Simulator.

After defibrillation of the manikin, the Simulator simulates defibrillation
artifact, followed by the pre-selected rhythm. The programmed post
defibrillation rhythm can be changed at any time.