Heart rate adjustment – Fluke Biomedical DataSim 6100 User Manual

Page 42

Advertising
background image

DataSim 6100

Operators Manual

2-6

Table 2-3. Calibration Functions (cont.)

Function

Adjusment

Levels

(78) HR CAL

(BPM)

(82) ECG/SIZE/CAL

(mV)

(84) RESP SIZE

CAL (

Ω

)

+8 203

+9 227

+10 250

+11 268

+12 300

To adjust calibration levels:

1. Using the keypad, press

SPEC FUNC

, followed by the numbers

78, 82,

or 84

. For example, using the number

78

, the keypad briefly displays:

ECG RATE

followed by the prompt:

], [ or STD

2. Press ] to increase or [

to decrease the value, or press

STD

to reset to the

standard level.

Heart Rate Adjustment

The Rate Adjust function, Special Function 53, offers the user the option of
adjusting rates up or down in 10% increments. Table 2-4. lists the heart rates
available for the various rhythms. On power-up, each rhythm defaults to the
standard heart rate (STD).