Fluke Biomedical DataSim 6100 User Manual
Page 20

DataSim 6100
Operators Manual
1-8
Table 1-2. Simulator Front Panel Controls and Indicators (cont.)
Label Component
Description
E
Aux Input/Output
Connector
Cable interconnects for optional
waveform, such as capnography and
IABP
F
Manual Defib Switch
Manually simulates defibrillation
G
Defib Connector
Input for Arrhythmia Anne or Chris
Clean manikins equipped with
defibrillation option
H
Personality Module
Connector
Input for optional modules
Used to add numerous functions and
waveforms to the standard set
I
Keypad Connector
Input for hand-help keypad
J
Keypad Rest
Storage area for the keypad
Inner
Well
Back Battery Charger
Input Connector
Jack used to recharge battery
The unit can be operated while
recharging.
Table 1-3. Lead ECG Amplitude
Lead #
ECG Amplitude
I
+.25 mv (+/- 5%)
II +1.0
III +.75
AVR .60
AVL -.25
AVF +.90
V1 .40
V2 .90
V3 +.60
V4 +1.0
V5 +1.5
V6 +
1.0