Operation

Defibrillation

Training

2

2-9

followed by the prompt:

CHAN?

2. Enter number

6

. The following prompt displays:

], [ or STD

Press

] t

o increase,

[

to decrease, or

STD

to reset to standard level.

Respiration amplitudes differ, depending upon which lead is monitored. Tables
2-2 and 2-3 list amplitudes for the RA-LA lead.

Note

For calibration applications, use the Respiration Size Cal function,
Special Function 84, to display impedance amplitudes. Pressure
waveforms are modulated by the respiration waveforms to simulate
respiration artifact. The respiration artifact level can be varied using
Pressure/Respiration Artifact, Special Function 67.

Cardiovascular artifact or ECG/Respiration coincidence can be
simulated by selecting the CVA Coincidence, Special Function 76.
The impedance signal is synchronized with the ECG. Impedance
decreases after each R wave, simulating a decrease in impedance due
to an increase in pulmonary blood volume during systole.

Defibrillation Training

The Defibrillation Training mode lets the user interface the Simulator with the
Laerdal Arrhythmia Anne and the Armstrong Chris Clean manikins for
realistic defibrillation and ECG paddle monitoring training.

To train in defibrillation using a manikin:

1. Follow the guidelines provided by the training manikin for proper cable

connection.

2. Plug the torso skin cable into the

DEFIB MANIKIN

jack.

3. Power on the Simulator.

4. On the keypad, select a life-threatening arrhythmia, such as

V-FIB