Fluke Biomedical DataSim 6100 User Manual

Page 27

Advertising
background image

Introduction and Specifications

Instrument

Familiarity

1

1-15

Table 1-5. Pendant Keypad Codes and Functions (cont.)

Code Function

Description

49 MUSCLE

ARTIFACT

Artifact amplitudes 0.4, 0.6 and 1.0 mV
peak to peak; use

]

and

[

keys or

eliminate with STD key.

50

CPR

CPR artifact a 70/min

ADJUST

53 ECG

RATE

ADJUST

Increases or decreases rate in 10%
increments; STD returns to standard
rate.

54

SIZE ADJUST

Increases and decreases ECG, BP, and
respiration waveforms or returns them to
standard size; use after selecting
desired channel.

55

AUTO TREND

Automatically varies NSR heart and
respiration rates and blood pressures;
Vtach episode 4 every 1 hour; trend
repeats every 2 hours.

DEFIB CONVERSATION

59 RHYTHM

SELECT

Permits selection of post defib rhythm,
which is then automatically inserted after
defib artifact.

60 DEFIB

DISABLE

Disables

auto defib sequence

RESPIRATION

71

APNEA

Thoracic impedance is simulated via
ECG output snaps.

72

RESP AT 10 BPM

Respiration waveform at 10 BPM

73

RESP AT 20 BPM

Respiration waveform at 20 BPM

74

RESP AT 40 BPM

Respiration waveform at 40 BPM

75

RESP AT 80 BPM

Respiration waveform at 80 BPM

76 CVA

COINCIDENCE

ECG and respiration are synchronized
and timed appropriately for
cardiovascular artifact. Impedance drops
immediately after each R wave.