Blood pressure cables – Fluke Biomedical DataSim 6100 User Manual

Page 55

Advertising
background image

Operation

Blood

Pressure

Cables

2

2-19

5. Set up the CO computer in accordance with the manufacturer instructions.

When ready to inject, select

SPEC FUNC

and the

appropriate

number

for the desired CO curve listed on the module.

* Note

The internal resistor value is selected to meet the input requirements
of the American Edwards or SpaceLabs bath probe injectate
temperature. Fluke suggests using the actual bath probe and bath to
meet the input requirements for other desired cardiac output
computer systems or modules. Some patient monitoring systems
default to an iced injectate (0 °C) if no bath probe is attached to the
system.

Blood Pressure Cables

To ensure proper operation, the interface cables used between Simulator
pressure outputs and the equipment must include the proper connectors and
wiring. For the appropriate cables, refer to Table 2-7.

Table 2-7. Appropriate Pressure Simulator Interface Cables

Model Item Model

Item

4100-01 Unterminated

4100-14

Abbott

(7m)

4100-02 HP

5

µ

V (5F)

4100-15

Gould/Care (5M)

4100-03 HP

40

µ

V (5F)

4100-18

Litton (4M)

4100-04 HP

5

µ

V (12M)

4100-19

Siemens (10M)

4100-05 HP

40

µ

V (12M)

4100-20

Kontron/Rouche (6M)

4100-06

Spacelabs (5M) (700,
Alpha 9, 14)

4100-23

Marquette (8M)

4100-08 GE

3100/Datascope

(6F)

4100-24

Burdick DataSim