Specifications – Fluke Biomedical DataSim 6100 User Manual

Page 31

Advertising
background image

Introduction and Specifications

Specifications

1

1-19

Specifications

Dimensions

........................................................... 13 in H x 10 in W x 4.7 in D

(33 cm H x 25.4 cm W x 11.94 cm D)

Weight

.................................................................... .05 lb (3.2 kg)

Power

AC power ............................................................ 100 V/240 V, 1.6 A max, 50-60 Hz;

output: +20V, 2.5 A; output power:
50 W max

Battery Type ....................................................... 12 V, 1.9 AH sealed lead acid

rechargeable

Battery Capacity ................................................. 20 hr
Battery Charge Time........................................... 5 to 95 % of complete charge in 14 hr

Temperature ..........................................................

10 ºC to 40 ºC

Humidity ................................................................

80% maximum

ECG

Output signals..................................................... High Level: 1 V/1 mV

Heart Rate

Range ................................................................. 30 to 300 BPM
Accuracy ............................................................. ±1 %

Output Connectors

Low Level............................................................ 12-Lead electrode snaps
High Level........................................................... Switchcraft 15GM7F
Pacemaker Artifact ............................................. -8 mV, 1 ms
Sample Rate (Maximum) .................................... 250 sample/s

Performance Testing

Linearity .............................................................. 2.5 Hz triangular wave
HR Cal Check ..................................................... 30 to 300 BPM
Chart Speed........................................................ 2.5 Hz square wave
Amplitude............................................................ 0.25, 0.50, 1.00, 1.50, 2.00, and

2.50 mV

Blood Pressure

Output signals

High Level........................................................... 1 V/100 mmHg
Transducer.......................................................... 5 and 40 µV/V mmHg
Exciter Voltage.................................................... 10 V ac or dc max
Static Range ....................................................... 0 to 250 mmHg, adjustable in 5 %

increments

Accuracy ............................................................. ±5 % or 1 mmHg, whichever is greater