Pac and pvc insertion – Fluke Biomedical DataSim 6100 User Manual

Page 29

Advertising
background image

Introduction and Specifications

Instrument

Familiarity

1

1-17

Table 1-5. Pendant Keypad Codes and Functions (cont.)

Code Function

Description

65

DAMPED LINE

Outputs a 50 mmHg flush waveform on
PA channel.

PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate damped frequency
characteristics of pressure line.

66 CATHETER

WHIP

Pressure artifact caused by motion of
catheter; on PA channel

67 PRES/RESP

ARTIFACT

Varies level of respiration artifact on
pressures

PROCEDURES

68

IABP

Arterial pressure with cardiac assists
pressure pulse, 2:1 augmentation.

Simulation only; not interactive with the
IABP; output on arterial channel

69

PA WEDGE

Outputs a PAW waveform

70 SWAN-GANZ

INSERTION

Output on PA channel: right atrium to
right ventricle; right ventricle to
pulmonary artery; pulmonary artery to
wedged pulmonary artery; wedged
pulmonary artery to pulmonary artery

PAC and PVC Insertion

The third row of the keypad has three palpitation simulation keys:

9/PAC

PVC1

PVC2

A single PAC or PVC can be inserted into any rhythm by pressing one of these
keys. The respective PAC or PVC is inserted into the next cardiac cycle. Only
one PAC, PVC1, or PVC2 can be inserted at any one time. The eight-character
LCD display responds to key entries by displaying the abbreviated function
name.