Table 1-5. Pendant Keypad Codes and Functions (cont.)

Code Function

Description

07 2nd

DEGREE

AV

BLOCK II

Slow regular rhythm with normal P
wave; PR interval in conduced beats is
normal; 3:1 block

08 3rd

DEGREE

AV

BLOCK

Slow regular ventricular rhythm; regular
atrial rhythm; independent atrial and
ventricular rhythms

09

PAC INSERT

Premature atrial contraction; a single
PAC is inserted into the current rhythm.

10

PVC1 INSERT

Premature ventricular contraction; a
Single PVC1 is inserted into the current.

11

PVC2 INSERT

Premature ventricular contraction; a
single PVC2 is inserted into the current.

12

COUPLET

Irregular rhythms with frequent couplets
and 1 pair of ventricular beats; PVCs are
premature; compensatory pause follows
couplet.

13 VENTRICULAR

TACHYCARDIA

Irregular, rapid rhythm with irregular,
successive PVCs; no P waves

14 VENTRICULAR

FIBRILLATION

Irregular, rapid, chaotic waveforms with
no QRS

ATRIAL

15 ATRIAL

TACHYCARDIA

Rapid, regular rhythm with biphasic P

16 ATRIAL

TACHYCARDIA/
ABERRANCY

Rapid, regular rhythms with biphasic P
wave and notched wide QRS.

17 FREQUENT

PACs

Irregular rhythm with frequent premature
PACs and a biphasic P wave

18 PEDIATRIC

TACHYCARDIA

Rapid regular rhythm with tall R wave
and narrow QRS

BLOCKS

19

1st DEGREE AV
BLOCK

Normal beats except with long PR
interval of .25 sec