Cardiac output personality modules – Fluke Biomedical DataSim 6100 User Manual

Page 54

Advertising
background image

DataSim 6100

Operators Manual

2-18

Cardiac Output Personality Modules

The Simulator can be used with either of two Cardiac Output (CO) Personality
Modules. 0 °C injectate temperature (model 6070-18) and 25 °C injectate
temperature (model 6070-20) both simulate the Edwards style blood
temperature catheter in generating several dilution curves.

The dilution curve at 0 °C injectate temperature assumes the following
conditions:

10 cc Injectate volume

Edwards 93-131-7F Catheter

Injectate Bath Probe, 90.22 k

Ω

(compatible with Switchcraft connector

SL-40-47F)

Computation Coefficient .542 *

The dilution curve at 25 °C injectate temperature assumes the following
conditions:

10 cc Injectate volume

Edwards 93-131-7F Catheter

Injectate Bath Probe 62.34 k

Ω

(compatible with Switchcraft connector

SL-40-47F)

Computation Coefficient .595 *

* Note

The use of other computation LM10 coefficients could result in CO
values other than specified.

To use the Cardiac Output Module:

1. Insert the Cardiac Output Module into the Simulator

Personality

Module

connector.

2. Power on the Simulator.

3. Connect the CO computer BT (blood temperature) cable to the CO

Module’s 3-pin connector. (The BT cable in the clinical setting is attached
to the thermistor on the Swan-Ganz catheter.)

4. Connect the CO computer injectate temperature cable to the black

injectate temperature box that simulates 0 °C or 25 °C temperature. *