DataSim 6100

Operators Manual

2-4

followed by the prompt:

CHAN?

2. Enter the desired channel number by pressing the numbers

1-6

. The

following prompt displays:

], [ or STD

3. Press ] to increase or [

to decrease the value, or press

STD

to reset to the

standard level. Table 2-2 shows the effect of incremental waveform
amplitude size adjustments on the parameters.

Table 2-2. Effects of Size Adjustments on Parameters

Channel/

Parameter

-3 -2 -1

STD

+1 +2 +3 +4

(1) ECG
R Wave
Amplitude (mV)

.25 .50 .75

1

mV

1.25

1.50

1.75 2.0

(2) ART
SYS/DIAS
Mean (mmHg)

116/62

80

123/66

85

130/70

90

137/73

95

143/76

100

150/80

105

157/84

110

164/88

115

(3) PA
SYS/DIAS
Mean (mmHg)

14/6

9

21/9

13

27/12

17

33/15

22

41/19

27/\

48/22

32

55/25

37

61/28

42

(4) RA Mean
(mmHg)

Not Adjustable

(5) AUX
(CO)

2 4 6 8 10 12 14 16

(6) RESP
(RA-LA) (

Ω

)

.25 .50 .75 1.0 1.25

1.50

1.75 2.0