Chan, Or std – Fluke Biomedical DataSim 6100 User Manual
Page 40

DataSim 6100
Operators Manual
2-4
followed by the prompt:
CHAN?
2. Enter the desired channel number by pressing the numbers
1-6
. The
following prompt displays:
], [ or STD
3. Press ] to increase or [
to decrease the value, or press
STD
to reset to the
standard level. Table 2-2 shows the effect of incremental waveform
amplitude size adjustments on the parameters.
Table 2-2. Effects of Size Adjustments on Parameters
Channel/
Parameter
-3 -2 -1
STD
+1 +2 +3 +4
(1) ECG
R Wave
Amplitude (mV)
.25 .50 .75
1
mV
1.25
1.50
1.75 2.0
(2) ART
SYS/DIAS
Mean (mmHg)
116/62
80
123/66
85
130/70
90
137/73
95
143/76
100
150/80
105
157/84
110
164/88
115
(3) PA
SYS/DIAS
Mean (mmHg)
14/6
9
21/9
13
27/12
17
33/15
22
41/19
27/\
48/22
32
55/25
37
61/28
42
(4) RA Mean
(mmHg)
Not Adjustable
(5) AUX
(CO)
2 4 6 8 10 12 14 16
(6) RESP
(RA-LA) (
Ω
)
.25 .50 .75 1.0 1.25
1.50
1.75 2.0