Operation

Calibration

2

2-5

Note

Pressure and R wave values are defined for normal sinus beats only.
To obtain non-varying pressure valves, select SPEC FUNC 67 and
decrease minimum level.

For calibration of performance check applications, use SPEC FUNC
78 (Heart Rate Cal), SPEC FUNC 81 (Linearity/Speed), SPEC FUNC
82 (ECG Size Cal), and SPEC FUNC 84 (Resp Size Cal).

Calibration

The Calibration functions, Special Functions 78, 82, and 84, let the user adjust
calibration levels for heart rate, linearity/speed, ECG size, and respiration size,
as illustrated in Table 2-3.

Table 2-3. Calibration Functions

Function

Adjusment

Levels

(78) HR CAL

(BPM)

(82) ECG/SIZE/CAL

(mV)

(84) RESP SIZE

CAL (

Ω

)

-4 30

-3 40

0.25

0.25

-2 50

0.50

0.50

-1 60

0.75

0.75

STD 70

1

1

+1 80

1.25

1.25

+2 90

1.5

1.5

+3 100

1.75

1.75

+4 121

2.0

2.0

+5 142

2.5

2.5

+6 163

+7 183