DataSim 6100

Operators Manual

1-16

Table 1-5. Pendant Keypad Codes and Functions (cont.)

Code Function

Description

77 PEDIATRIC

APNEA/BRADY

Intermittent apnea periods; heart rate
drops from 140 BPM to 70 BPM each
apnea period.

CALIBRATION

78 HEART

RATE

CALIBRATION

Calibrated heart rates from 30 to 300
BPM

81 LINEARITY/SPEED

2.5

Hz

triangular ECG waveform

82 ECG

SIZE

CALIBRATION

Cal pulse amplitudes from .25 mV to
2.5 mV

83 ARRHYTHMIA

SEQUENCE

Automatic sequence of arrhythmias
over 3 minute period of time: NSR,
Dropped Beat, AV Pace, V Pace,
Atrial Tach, Couplet, RBBB, Triplet,
Bigeminy (3 Foci), Vtach

84 RESPIRATION

SIZE CAL

Respiratory depth ranging from 0.25 to
2.5 ohm; calibrated for RA-LA lead

BLOOD PRESSURE

62 ZERO

PRESSURE

Zeros all three-pressure channels as
well as auxiliary CO2 channel

63

PATENT LINE

Outputs a 50 mmHg flush waveform on
PA channel

PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate ideal frequency
characteristics of pressure line.

64

RESONANT LINE

Outputs a 50 mmHg flush waveform on
PA channel

PA waveform follows flush. Trailing edge
of flush and PA pressure waveforms
demonstrate resonant frequency
characteristics of pressure line.