Installation and initial start-up – Ivoclar Vivadent EP 5010 User Manual

Page 19

Advertising
background image

19

4.4 Removing the furnace head

Before the connection cover and the covering hood are removed, the
furnace has to be switched off and the power cord disconnected
from the power socket.

1. Loosen and remove the knurled screw of the connection cover.
2. Dismount the connection cover.
3. Open the covering hood.
4. Release the heater plug and disconnect it with a 45° anti-

clockwise rotation.

5. Disconnect the press drive plug.
6. Disconnect the thermocouple plug.
7. Disconnect the vacuum hose.
8. Press the leaf spring with a finger, lift off the furnace head at the

same time and remove it.

Make sure that the furnace head has completely
cooled down before it is removed (fire hazard).

4.5 Initial start-up

1. Connect the power cord with the wall socket.
2. Switch the On / Off button at the rear of the furnace to position I.

4.5.1 Basic settings upon initial start-up
Upon the initial start-up of the new furnace, a number of basic
settings are required. These settings will be stored and will not appear
anymore upon the following starting procedure.

Step 1:
Select the language.
The touch buttons (display keys) can be operated by tapping the display.

Step 1

Step 2

Step 3

Select the desired language using the [Arrow up / down] buttons. Confirm the entry with the green button. The [Next]
button is used to reach the next entry screen.

4. Installation and Initial Start-Up

Advertising