Operation and configuration – Ivoclar Vivadent EP 5010 User Manual

Page 57

Advertising
background image

57

2. Conduct the software update.
If the USB stick with the software file is already connected, the
furnace automatically searches for a valid software file. If the USB
stick has not yet been connected with the furnace, do so now.
Press the [Execute] button.

3. The status bar shows the progress of the update.

4. Complete the software update.
The following message is displayed:

Software update successful

Software update failed

The furnace has to be switched off and on again by means of the mains switch at the rear of the furnace to

complete the software update.

Important information

Please note that modified Ivoclar Vivadent programs may be overwritten during a software update. Individual
programs are not affected and will not be overwritten.

5.4.5 Software update
Software updates can be easily installed on the furnace by means of a USB stick. A USB stick which contains the current
software file (e.g. EP 5010_V1.10.iv) is required for the update. The software version on the USB stick must be higher than
the one installed on the furnace (see Selection – Information).

The free software updates for Programat furnaces are available from www.ivoclarvivadent.com/downloadcenter.

A software update requires only few steps:

1. Open the Software Update menu.
Scroll to page 4 in the home screen and press the [Software Update]
button.