7. Maintenance, Cleaning and Diagnosis

This chapter describes the user maintenance and cleaning procedures for the Programat EP 5010. Only those tasks are
listed that may be performed by dental professionals. All other tasks must be performed by qualified service personnel at
a certified Ivoclar Vivadent Service Center.

7.1 Monitoring and maintenance

The time for these maintenance procedures depends on the frequency of use and the working habits of the users.
Therefore, the recommended times are only approximate values.

Hazardous area

Part

When

Check all plug-in connections for correct fit.

Various external connections

Weekly

Check if the furnace head opens smoothly and without excessive
noise.

Furnace head opening mechanism Monthly

Check if the thermocouple is straight and in the right place.

Thermocouple

Weekly

Check the insulation for cracks and damages. If the insulation is
worn down, it has to be replaced by a certified Ivoclar Vivadent
Service Center. Fine hairline cracks on the surface of the insulation
are harmless and do not influence the function of the furnace in a
negative fashion.

Insulation

Monthly

Check if the sealing rims of the furnace head and furnace base are
clean and undamaged.

Sealing rims of the furnace head
and base

Weekly

Check the keypad for visible damage. If the keypad is damaged, it
has to be replaced by a certified Ivoclar Service Center.

Membrane-sealed keypad

Weekly

Check the temperature:
Use the temperature checking set to check and adjust the tem-
perature in the furnace.

Firing chamber

Twice a year

Check the quartz glass cylinder to make sure that the quartz glass
is not defective.

Firing chamber

Daily

Check if there is condensate in the vacuum hose or firing chamber.

Vacuum hose,
firing chamber

Monthly

This furnace has been developed for typical use in dental laboratories. If the product is used in a production

enterprise, for industrial applications and for continuous use, premature ageing of the expendable parts has to
be expected. The expendable parts are, e.g. the heating muffle, the furnace head sealing rim or the insulation
material.

Expandable parts are not covered by the warranty. Please also observe the shorter service and maintenance intervals.

In general, the furnace head should not be replaced since the components (furnace head and furnace base) have

been coordinated with each other. However, if the furnace head must be replaced for maintenance reasons,
subsequent temperature calibration is required.