5.4.7 Maintenance programs
Scroll to page 3 in the home screen and press the [Maintenance
Programs]
button.

The following programs are available in the Maintenance Programs
menu:
– dehumidification program
– cleaning program.

2. Start the dehumidification program.
Press the START key to start the dehumidification program.

5.4.7.1 Dehumidification program
Condensation in the insulation of the firing chamber or the vacuum pump results in a poor vacuum and thus in poor firing
results. For that reason, the furnace head should be kept closed when the furnace is switched off or is below 100 °C in
order to prevent the absorption of humidity.

Conducting the dehumidification program:

1. Open the dehumidification program.
Scroll to page 3 in the home screen and press the [Maintenance
Programs]
button. Press the [Dehumidification Program] button in
the Maintenance Programs menu.