Operation and configuration – Ivoclar Vivadent EP 5010 User Manual

Page 55

Advertising
background image

55

6. Start the calibration.
Press the START key to start the calibration program. The progress of
the calibration program is shown in the display.

7. Complete the calibration.
The result is displayed at the end of the calibration.

Temperature calibration successful

Temperature calibration failed

The calibration value is the difference between the measured current
temperature and the desired set temperature.

At the end of the program, open the furnace head and carefully remove the ATK2 sample using the furnace tongs and
place it on the cooling tray to allow it to cool. Reinsert the firing plate using the furnace tongs.

To return to the home screen, press either the [Home] touch button in the navigation bar or the HOME key on

the membrane-sealed keypad.

5. Operation and Configuration

5. Press the sample into place.
If necessary, use the furnace tongs to apply slight pressure to the
centre of the calibration base until the calibration sample clicks into
place. Observe the corresponding markings.