Operation and configuration – Ivoclar Vivadent EP 5010 User Manual

Page 54

Advertising
background image

54

2. Start the calibration.
The display shows the last calibration value and the date of the
last calibration.

Press the START key on the membrane-sealed keypad to start the
calibration. Follow the instructions on the display.

5.4.3 Temperature calibration
The thermocouple and heating muffle of the furnace may be subject to changes which affect the furnace temperature,
depending on the mode of operation and the frequency of use. Conduct an automatic temperature calibration at least
every six months.

Temperature calibration requires only few steps:

1. Open the temperature calibration.
Scroll to page 2 in the home screen and press the [Temperature
Calibration]
button.

The furnace must be heated up and have reached its

stand-by temperature (403 °C) before the calibration is
started.

3. Remove the firing plate.
Remove the firing plate from the furnace using the furnace tongs
and place it on the cooling tray.

4. Insert the ATK2 sample.
Carefully grip the upper part of the ATK2 using the furnace tongs
(Caution: fracture risk of the ceramic) and insert it into the holes
designated for this purpose until it snaps into place.

5. Operation and Configuration