Checking pre-installed software
Software is pre-installed on K2 products when you receive them from the factory. This load of
pre-installed software is referred to as the “golden drive”. The following list is an example of the
software pre-installed. Check "K2 Release Notes" for the most up-to-date list with version
information.
If you suspect that pre-installed software is not correct, use the recovery process to re-load the
software. Do not attempt to un-install, install, or repair pre-installed software without guidance from
your Grass Valley Support representative.
K2 Solo 3G system pre-installed software
•
Intel Pro Software
•
QuickTime
•
Microsoft iSCSI Initiator
•
MS XML
•
.NET Framework
•
MegaRAID — Do not use this utility on a K2 Solo 3G system. This utility is for use by qualified
Grass Valley Service personnel only. When this utility is opened it scans the SCSI bus and
interferes with record and play operations.
•
J2SE Runtime Environment
•
StorNext software
•
Windows PowerShell
•
Windows XP Embedded
Making CMOS settings
NOTE: This procedure is intended for use by Grass Valley Service personnel or under the direct
supervision of Grass Valley Service personnel.
1. Connect keyboard, monitor, and mouse to the K2 Solo 3G system.
2. Restart the K2 Solo 3G system.
3. During the BIOS startup screen, watch the keyboard lights (capslock, numlock, etc.). When the
lights flash, press
Delete
to enter Setup.
4. Press
F3
and then press
Enter
. This loads optimal default values for all the setup questions.
5. Press
F4
and then press
Enter
to save settings and restart.
Restoring disk controller configuration
Do this task when replacing the disk controller board.
This task can be used on any K2 Solo 3G system, but it is required on any system that has a Type
II (ADLINK) CPU carrier module. This includes the first generation K2 Summit system, which can
have a Type II CPU carrier module that was installed in the factory or that was upgraded in the field.
56
K2 Solo 3G Service Manual
06 November 2012
Service procedures