5 set up delay time, Set up delay time, See section 9.3.4 ) – Liebert Universal Monitor User Manual

Page 55

Advertising
background image

System and Control Options

47

9.3.4 Set Up Alarmable Inputs in Latched or Unlatched Mode

Alarmable input points may be set up in Latched mode (Y), which

requires the user to clear the Universal Monitor alarms after an alarm

has occurred, or Unlatched mode (N), in which alarms will automati-

cally clear after a return-to-normal state. The default setting for

alarmable inputs is N (Unlatched). Any input defined as an event is

automatically Unlatched.
A digital input defined as Latched does not reset automatically after

an alarm. It will remain in the triggered position even after the moni-

tored device returns to normal. After an alarm, the user must manu-

ally reset Latched inputs, as described in 9.12.1 - Clear Active

Alarms. This also means that any digital outputs linked to a Latched

digital input will remain triggered until the input is manually reset.
A digital input defined as Unlatched will automatically track the sta-

tus of the monitored device. No action by the user is required.

• From the Setup Input Menu, choose Latched, as shown above

right, and press Enter ↵ .

• To change this feature, use the arrows

↑ ↓

to

choose Y (Yes -

Latched) or N (No - Unlatched), then press Enter ↵ .

NOTE: Only alarmable input points may be set up as Latched. If you

attempt to set up an event as Latched, the message shown at right

appears. Press the ESC key to return to the previous screen to con-

tinue configuring the digital input.
The only way to reset latched alarms is by clearing active alarms. This

step is required to return latched points to their normal state. For

details, see 9.12.1 - Clear Active Alarms.

9.3.5 Set Up Delay Time

Each input may be set up with a specified time delay between the

occurrence of an alarm or event—a monitored device changes state—

and any response by the Universal Monitor. This feature can prevent

needless responses to transient conditions.
This delay is the amount of time that must elapse before the Universal

Monitor acknowledges the change of state in the monitored device and

takes any further action, such as dialing configured pagers. The

default time delay is 1 second (displayed as 00 MIN : 01 SEC).
The time delay applies only when a monitored device changes from its

normal state—for example, the Universal Monitor detects a closed

contact that is defined as Normally Open.
When the monitored device returns to its normal state, the Universal

Monitor acknowledges the change of state immediately.

• From the Setup Input Menu, choose Delay, as shown at right, and

press Enter ↵ .

• To change the time delay in minutes and seconds (in the range of 00:00 to 99:59):

• For minutes (MIN), press and hold one of the arrow keys

↑ ↓

scroll through the numbers

00-99.

• Press Enter ↵ to advance to seconds (SEC).
• For seconds (SEC), press and hold one of the arrow keys

↑ ↓

scroll through the numbers 00-59.

• When finished, press Enter ↵ again.

Setup Input Menu

SETUP INPUT_01

LABEL: INPUT_01
NORM OPEN/CLOSE: NO
ALARM/EVENT: AL

↑ ↓=NEXT

=SELECT

> LATCHED: N

DELAY: 00:01

Latched / Unlatched

IS ALARM
LATCHED? N

↑ ↓=Y/N

=END

Message (if input has

been defined as an event)

LATCH NOT ALLOWED
WHEN IN EVENT MODE

ESC=BACK

Setup Input Menu

SETUP INPUT_01

LABEL: INPUT_01
NORM OPEN/CLOSE: NO
ALARM/EVENT: AL

↑ ↓=NEXT

=SELECT

LATCHED: N
DELAY: 00:01

Delay Time

DELAY TIME BEFORE
TRIGGER?

00 MIN : 01 SEC

↑ ↓=SELECT

=NEXT/END

Advertising