Get Assistance, if You Need It

6

Agilent E6474A User’s Guide

111

Accuracy

The accuracy is calculated as:

Frequency accuracy = signal generator accuracy + 1 ppm.

Amplitude accuracy = ± (0.5 dB + signal generator accuracy +
mismatch error + cable loss).

Assuming the Agilent receiver and Agilent 8648B signal
generator specifications, with nominal values of mismatch and
cable losses of approximately 1 dB, you can expect:

Frequency reading, approximately: Frequency selected ±450

kHz

Amplitude reading, approximately: Amplitude selected ±2.5

dB

Passing the test

If the display shows a single spectral peak, and reports
frequency and power levels that match the output of the signal
generator, the system has passed the test.

If your system is equipped with a navigator, perform the tests in

“Test 4 - GPS receiver testing"

on page 116. If you experience

problems in setting up the tests, refer to

“Troubleshooting your

E6474A system"

on page 102.

NOTE

You can save the settings of the project (.spf file) at any time. If you
experience measurement or configuration problems, you can refer to the
settings shown in the corresponding verification test.