ACQUISITION - INTERLEAVED, ILVD

The INTERLEAVED command enables or disables Equivalent Time Mode for timebase settings where

both single shot and RIS mode are available.
Equivalent Time Mode is not available for sequence mode acquisitions. If sequence mode is on, ILVD

ON turns it off.
The response to the INTERLEAVED? query indicates whether the oscilloscope is in Equivalent Time

Mode.

Command Syntax

InterLeaVeD <mode>

<mode>:= {ON, OFF}

Query Syntax

InterLeaVeD?

Response Format

InterLeaVeD <mode>

Example

The following instructs the oscilloscope to use Equivalent Time Mode:

ILVD ON

Related Commands

TIME_DIV, TRIG_MODE