Status - cmr, Query syntax, Response format – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 63: Related commands, Cmr? status register structure, Operator's manual

Advertising
background image

Operator's Manual

STATUS - CMR?

The CMR? query reads and clears the contents of the CoMmand error Register (refer to the following

table for details) which specifies the last syntax error type detected by your oscilloscope.

Query Syntax

CMR?

Response Format

CMR <value>

<value>:= 0 to 13

Related Commands

ALL_STATUS?, *CLS

CMR? Status Register Structure

Value Description

1

Unrecognized command/query header.

2

Illegal header path.

3

Illegal number.

4

Illegal number suffix.

5

Unrecognized keyword.

6

String error.

7

GET embedded in another message.

10

Arbitrary data block expected.

11

Non-digit character in byte count field of arbitrary data block.

12

EOI detected during definite length data block transfer.

13

Extra bytes detected during definite length data block transfer.