Acquisition - volt_div, vdiv, Command syntax, Query syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 35: Response format, Example

Advertising
background image

Operator's Manual

ACQUISITION - VOLT_DIV, VDIV

The VOLT_DIV command sets the vertical sensitivity in Volts/div.
The probe attenuation factor is not taken into account for adjusting vertical sensitivity.
The VOLT_DIV? query returns the vertical sensitivity of the specified channel.

Command Syntax

<channel>:Volt_DIV <v_gain>

<channel>:= {C1, C2, C3, C4}

<v_gain>:=

Refer to product datasheet at teledynelecroy.com.

NOTE: When specifying <v_gain>, the unit V is optional.

Query Syntax

<channel>:Volt_DIV?

Response Format

<channel>:Volt_DIV <v_gain>

Example

The following instruction sets the vertical sensitivity of channel 1 to 50 mV/div:

C1:VDIV 50MV

920836 Rev B