Program messages, Command/query syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 8

Advertising
background image

WaveAce Remote Control

2. From the WaveAce menu, choose Utility > IP Settings (on the third page of settings).

3. Enter an IP Address, Subnet Mask, and Gate Way. Enter values by selecting each segment and

turning the Adjust knob until the desired number is reached. Press the top soft key to jump to the

next field or press the Adjust knob to tab to the next segment of the current field. Leave DHCP

disabled.

Program Messages

Program messages are composed of commands or queries separated by

semicolons

and ending with a

terminator

:

<command/query>; . . . . . ;<command/query> <terminator>

Command/Query Syntax

The general form of a command or a query is an optional

header path

, followed by a command

header

,

optionally followed by one or several

parameters

(shown as

<data>

in the following construct):

[header_path:]<header>[?] [<data>,...,<data>]

NOTE:

There is a space between the header and the first parameter.

Commas separate parameters.

The question mark [?] is optional and turns the command into a query.

The oscilloscope does not distinguish between upper- and lowercase characters. Commands/queries are shown

mixed-case in this manual to call out the short form within the long form.

6

920836 Rev B

Advertising