ACQUISITION - WAIT

The WAIT command prevents your instrument from analyzing new commands until the current

acquisition has been completed. The optional argument specifies the timeout (in seconds) after which

the scope stops waiting for new acquisitions. If <t> is not given, or if <t> = 0.0, the scope waits

indefinitely.

Command Syntax

WAIT [<t>]

<t>:=

timeout in seconds (default is indefinite)

Example

ARM;WAIT;*OPC;C1:PAVA? MAX

This example finds the maximum amplitudes of several signals acquired one after another. ARM starts

a new data acquisition. The WAIT command ensures that the maximum is evaluated for the newly

acquired waveform.
C1:PAVA? MAX instructs the oscilloscope to evaluate the maximum data value in the Channel 1

waveform.

Related Commands

*TRG, TRIG_MODE, ARM