ACQUISITION – PEAK DETECT, PDET

The PEAK_DETECT command switches ON or OFF the peak detector built into the acquisition system.
The PEAK_DETECT? query returns the current status of the peak detector.

Command Syntax

Peak_DETect <state>

<state> : = {ON, OFF}

Query Syntax

Peak_DETect?

Response Format

PDET <state>

Example

The following instruction turns on the peak detector:

PDET ON

ACQUISITION - STOP

The STOP command immediately stops the acquisition of a signal. If the trigger mode is AUTO or

NORM, STOP places the oscilloscope in STOPPED trigger mode to prevent further acquisition.

Command Syntax

STOP

Related Commands

ARM_ACQUISITION, TRIG_MODE, WAIT, FORCE_TRIGGER